Scheduling & Maintenance
Dependable scheduling for all of our services
Dependable scheduling for all of our services
Meet all NSF/ANSI 49 Standards
Full inventory of lab equipment.
AIE Lab Services and American Instrument Exchange have been working with biosafety cabinets and clean air systems from The Baker Company, Nuaire, Thermo Scientific, Labconco, Kewaunee, Air Clean, Erlab, Flow Sciences, and more for over fifty years. Our staff has decades of combined experience with repairs, refurbishments, filter changes, and decontaminations, meaning we have a wealth of knowledge with us for every project we tackle.
All biological safety cabinets are certified to meet NSF-49 guidelines in order to verify that the cabinet provides protection to your product, person, and environment
The following are the potential tests performed as described by NSF:
Pressure Decay / Soap Bubble
Determines the overall seal integrity of the cabinet outer hull, including exterior surfaces of all plenums, welds, gaskets, plenum penetrations, and seals.
HEPA/ULPA Flter Lak Test
Determines the integrity of downflow and exhaust HEPA/ULPA filters, filter housings, and filter mounting frames.
Site Installation Assessment Tests
Performed to verify that the biosafety cabinet is integrated properly into the facility.
This includes:
Electrical Leakage and Ground Crcuit Resistance and Polarity Tests
All new cabinets shall conform the requirements of the current edition of any national standard that is based on IEC610-1. Cabinets initially qualified under versions of NSF/ANSI 49 prior to the 2009 edition shall conform to UL61010A-117 or may refer to NSF 49-1992 for Electrical Leakage, Ground Circuit Resistance, and Polarity tests if necessary.
Motor/Blower Performance
Demonstrates that when the cabinet is operated at the nominal set point velocities and without readjusting the fan speed control, a 50% increase in pressure drop across the new filter shall not decrease total air delivery more than 10%.
Downflow Velocity Test
Measures the velocity of air moving through the cabinet workspace 4 inches (100 mm) above the bottom edge of the sash.
Inflow Velocity (Face Velocity) Test
Determines the measured and calculated inflow velocity through the work access opening and the calculated exhaust flow volume rates.
Personal, Product, and Cross-Contamination Protection (Biological) Tests
Determine whether aerosols will be contained within the cabinet, outside contaminants will not enter the cabinet work area, and aerosol contamination of other equipment in the cabinet will be minimized.
Airflow Smoke Patterns Test
Determines the airflow along the entire perimeter of the work access opening is inward, that airflow within the work area is downward with no dead spots or refluxing, that ambient air does not pass on or over the work surface, and that there is no escape to the outside of the cabinet at the sides and top of the sash.
All biosafety cabinets certifications are performed with standard operating procedures which are updated regularly to comply with federal guidelines.
Due to the specialized knowledge required for a proper and safe certification of biosafety cabinets, NSF International administers an accreditation program to accredit field certifiers. Field certifiers must pass a written and practical test administered by NSF before becoming accredited. In addition, continuing education and periodic reexaminations are required for accreditation.
Velocity Profile Testing
Vertical and/or Horizontal Test in Accordance With:
Functional Testing
Acceptance Criteria (FPM)
Test Performance in Accordance
Test Performance
Test in Accordance with:
Supply Velocity Test
Airflow Smoke Pattern Test
Functional Test Result
Final Test Result
Duct Type
General Notes and Summary
Type of Protection
Personnel, product, and environment
Instrument Calibration Report
Only Per Client Request
Test in Accordance with:
Face Velocity Test
Airflow Smoke Pattern Test
If applicable —
HEPA /ULPA Filter Test – Filtration Mechanism
If applicable —
Functional Test Result
Final Test Result
Duct Type
Ducted or Non-ducted
General Notes and Summary
Type of Protection
Personnel, product, and environment
Instrument Calibration Report
Only Per Client Request
American Instrument Exchange buys, refurbishes, and sells high-quality used laboratory equipment for many types of labs, including biotech, biomedical, medical research, chemical, pharmaceutical, and university and hospital research labs.